Emcyte Corporation

 

 

 

BioSciences Research Associates, Inc.


 

1.3.2 Post-Process Platelet Yields

Table II provides summary data for platelet recovery in processed samples.

Table II: Post –Process Platelet Yields

Mean ± 1SD (Range)

PRP volume 4 ml

 

PRP @ 0hr

PRP @ 4hr

Yield (%)

68 ±17 (48-106)

65 ± 16 (44-96)

 

 

1.3.3 Platelet concentration (times baseline) in time 0 and Time +4 hr CPP

Table IV lists the calculated increase in platelet concentration in CPP samples above baseline in both Time 0 and 4 hr post-processing samples.

Table III: Post –Process Platelet Concentration

Mean ± 1SD (Range)

PRP volume 4 ml

 

PRP @ 0hr platelet

(Times Baseline)

PRP @ 4hr

(Times Baseline)

Correlation Coefficient

10 ± 3 (5.8-14)

9 ± 3 (6.6-14)

 


Post –Process Bone Marrow Aspirate
Mean ± 1SD (Range)
Bone Marrow Concentrate (BMC) volume 4 ml

 

 

Volume

 

Total Nuclear

Cells x 103/µL

Concentration

Above Baseline

Bone Marrow Aspirate Baseline

60

16-23

1.0X

Bone Marrow Concentrate

4

170-271

11.5X

 

 

Table I: Cell concentrations in 60 ml bone marrow aspirate and in 4ml bone marrow concentrate sample: Total nucleated cells (TNC), Platelets (Plt) and CD34+ hematopoietic stem cells (HSC), n=4

Bone Marrow Aspirate Baseline

Bone Marrow Concentrate

TNC x 106

HSC x 106

Plt x 106

TNC x 106

HSC x 106

Plt x 106

1161

± 239

9.0

± 1.3

10,830

± 1836

894

± 232

6.9

± 1.4

7,623

± 1432

Concentration factor (times baseline)

In 4ml buffy coat BMC 

11.6

11.5

10.6

 

 

The recovery of TNC, Plt and CD34+ HCS were calculated and are shown in Figure 1.

 Figure 1. Recovery of TNC, Plt and CD34+ HCS

 

 

Aspirate Volume (mL)

TNC In BAC x106

TNC x106

Per 60mL Aspirate

GenesisCS 60 896 896

Harvest BMAC

60 826 826

Gambro 2991

300 1170 358
Ficoll 60 373 373

 

 